did a calibration on the D-Dimer assay today. Despite following the
instructions things didn’t go according to plan. After a little struggling it
turns out that although the device can store multiple reagent lot data, it only stores one calibration file. When the
calibrator expiry date is reached the calibration and assay is automatically
invalidated even though the calibration is still acceptable.
Therefore
DO NOT have multiple calibrator information stored on-board.
I always say a day
when you learn nothing is a day wasted…